Health Products

ORT (2021-09-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Deerfield Imaging, Inc.
5101 Shady Oak Road
Minnetonka
55343
Minnesota
UNITED STATES
ORT300114093-600

Introduction

Affected Products

ORT300

Reason

The ORT200/300 operating room table is specifically designed to provide an mr compatible platform for neurosurgical use for cranial and spinal surgeries. The tables are removable from the operating room and mate to a docking station installed in the floor of the or via a floor latch mechanism which contains 4 linear actuators. Complaints have reported an inability to dock or undock the operating room table from the floor baseplate, due to malfunction of the actuators within the floor latch locking mechanism.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime