Health Products

Cardiosave System - Battery Only (2021-09-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
181725430IPCardiosave System - Battery Only
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
0998-00-0800-XX

Introduction

Affected Products

Cardiosave System - Battery Only

Reason

A cardiosave Lithium-ION battery pack (0146-00-0097) used during protocol testing failed to meet the minimum runtime requirement per getinge's internal product specification. These nonconforming batteries were inadvertently released to customers.