Health Products

Vidas products (2021-09-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.H.  Vidas HIGH SENSITIVE TROPONIN I (TNHS)
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
415386
1008693930F.  Vidas SYSTEM - Vidas NT-PROBNP2
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30 458
1008597480G.  Vidas C. DIFFICILE GDH
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30 125
More than 10 numbers, contact manufacturer.D.  Vidas - B-R-A-H-M-S PCT (PROCALCITONIN) ASSAY
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30 450
1008589450E.  Vidas - D-DIMER EXCLUSION II
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30 455
More than 10 numbers, contact manufacturer.B.  Vidas - PROGESTERONE ASSAY
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30409
1008731010C.  Vidas - C.DIFFICILE TOXIN A & B
Manufacturer
BioMérieux S.A
376 Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
30 118
More than 10 numbers, contact manufacturer.A.  Vidas - HCG ASSAY
Manufacturer
bioMérieux SA
376 Chemin de l'Orme
Marcy-l'Étoile
69280
FRANCE
30405

Introduction

Affected Products
  1. Vidas - HCG ASSAY
  2. Vidas - PROGESTERONE ASSAY
  3. Vidas - C.DIFFICILE TOXIN A & B
  4. Vidas - B-R-A-H-M-S PCT (PROCALCITONIN) ASSAY
  5. Vidas - D-DIMER EXCLUSION II
  6. Vidas SYSTEM - Vidas NT-PROBNP2
  7. Vidas C. DIFFICILE GDH
  8. Vidas HIGH SENSITIVE TROPONIN I (TNHS)

Reason

Based on an unusual rate of complaints from the field for "substrate error" leading to a potential delayed results on different Vidas References tested on all Vidas? Systems, Biomerieux initiated an investigation. The substrate error issue was confirmed on all lots of Vidas? Immuno-Assays manufactured with substrate batches using the common lot of raw material (4-MUP). The only potential risk associated with the substrate error is a delayed result as you need to run another test. There is no risk of false results. The investigation demonstrates that the issue occurs over the product shelf-life. The analyzes of the kinetic model allowed to define a revised expiry date for each impacted lot of Vidas? Immuno-Assays finished products. When used until the revised expiry date, the product continues to perform per its registered performance specifications